Test Connections
 

The GTX2200 and GC2200 series timer/counters allow the user to set a gate time through either the software front panel or through an API call GxCntSetGateTime().

The gate will stay open until one full period of the input signal has completed.

If the gate time is set to a value that is less than one period of the input signal, then the gate will actually remain open until one signal period has completed. This is why a valid measurement is still obtained even when the gate time is set to a value less than the input signal period. This behavior also explains the reason for the "plus up to one signal period" in the gate time specification (GTX2200 Datasheet). If the gate time expires in between a full period of the input signal, the gate will remain open until that cycle (period) completes.

An example of this is as follows (see figure below):
If the gate time is set to 500 uS, and the input signal's period is 200 uS, the gate will actually close at 600 uS, after the 3rd cycle has completed (it will not close mid-cycle at 500 uS). The function of the gate time allows the user to force the instrument to oversample an input signal. The instrument does not allow the user to select a gate time that will result in a closure part way through a cycle of the measured signal.

Timer/Counter

The Measurement Timeout function can be set by using the API function call GxCntSetMeasurementTimeout() to ensure that a very slow signal will not cause unnecessary delays in your program due to slow input signals or a signal that becomes intermittent or stops in mid period.
(If the Print button is not supported, simply use your
Web browser's "Print" command instead.)

Geotest-Marvin Test Systems, Inc.
1770 Kettering
Irvine, CA 92614
1-949-263-2222
1-888-837-8297

Geotest - Marvin Test Systems, Inc.
©2011 Geotest-Marvin Test Systems, Inc. (www.geotestinc.com). All rights reserved.
Geotest - Marvin Test Systems, Inc. - Geat A Head Start With Geotest