CalEasy supports the following calibration run modes:
Verify – in this mode no calibration is done. The instrument is only verified to its original specification and the result is printed to the test log.
Verify and Calibrate If Required – in this mode verification is performed and if required (verification task fails) the instrument is calibrated. If Calibration is successful, the instrument’s non-volatile memory is modified with the new calibration values and a calibration certificate is generated. If verification is OK, the instrument’s non-volatile memory is modified with the new calibration date (no new calibration values are written to the instrument non-volatile memory) and a certificate is generated.
Calibrate Always – in this mode no initial verification is performed. The calibration task is preformed and the certificate is generated upon successful calibration.