The following are specifications for GX5055 board:
Feature |
Value |
Timing |
|
Internal Test Clock |
5Hz to 50MHz |
Test Clock Timebase |
PXI 10 MHz clock |
Test Clock Resolution, max |
Greater of 1Hz or 0.01% of the programmed test clock frequency |
Programmable Data Sense Strobe |
0 to 63.75 ns, relative to drive clock edge (0 ns delay setting)250 ps resolution |
Channel I/O ( All specifications based on pin electronic voltage rails (VCC and VEE) of +18 V and -14 V) |
|
Data Direction control |
Per step, per channel |
Channels per board |
32 |
Number of drive and sense voltage references |
32 Drive Hi / Drive Lo, 32 Sense Hi / Sense Lo |
Drive voltage levels for specified rails (+18, -14 V) |
Drive Hi: -9 to +15 voltsDrive Lo: -10 to +11 volts |
Max Drive Level Range |
14 V to +26 V |
Drive voltage p-p range |
Min: 0.5 V p-p, Max: 25 V p-p |
Drive voltage accuracy |
+/- 25 mV, < 25 V p-p drive voltage |
Drive voltage resolution |
16 bits |
Driver leakage current |
+/- 25 nA max. ( Hi-Z mode) |
Output impedance |
50 ohms, typical |
Drive current |
200 mA per channel, 1.6 A per board, max. |
Short circuit protection |
Programmable current level with automatic disable, per channel basis |
Slew rate |
.1 to 1 V/ns, adjustable, programmable on a per channel basis |
Channel skew |
160 ps, typical, between the same card320 ps max., after calibration, for all channels ( Drive and Sense) |
Data Formats |
No return, Return to 1, Return to 0, Return to Hi-Z, Return to complement, Surround by complement; selectable on a per channel basis |
Drive Data timing |
Data assert on rising edge of test clockClock output to data valid:Data de-asserts on falling edge of test clock (for formatted data) |
Sense voltage range for specified rails |
Sense Hi: -10 to +11 VSense Lo: -10 to +11 V |
Max Sense Level Range |
-14 V to +22 V |
Sense voltage threshold accuracy |
+/- 25 mV, < 25 V p-p sense voltage |
Sense voltage resolution |
16 bits |
Input leakage current |
+/- 15 nA, max |
Minimum data sense pulse width |
4 ns, typical |
Data Sense timing |
Acquire data relative to ext strobe output clock edge |
Pull-up / Pull-down current source/ sink |
+/ -24 ma, programmable on a per channel basisV commutate: -10 to +11 V., programmable on a per channel basis |
Pull-up / Pull-down current source / sink accuracy |
+/- 250 uA |
Pull-up / Pull-down current source / sink resolution |
16 bits |
Voltage commutation accuracy |
+/- 25 mV, < 25 V range |
Voltage commutation resolution |
16 bits |
Resistive load |
Range: Hi-Z, 250 ohm, 1 K ohm, programmable on a per channel basis |
Memory depth (steps) |
512Kb per channel |
Sequencer and Trigger Functions |
|
Commands |
No Op, Set Reg, Loop, Call Subroutine, Return, Pause, Halt |
Trigger |
Software generated triggerExternal input trigger override |
Pause |
Software generated pauseExternal input pause overrideSequencer pause command |
External Timing, Control and Status Signals |
|
Clock Output (OClk) |
Data clock, TTL compatible;Programmable delay relative to internal system clock, 0-63.75 ns, 250 ps of resolutionJitter: 40 ps RMS |
External Strobe Output (Obst) |
Sense data strobe, TTL compatibleProgrammable delay relative to internal system clock, 0-63.75 ns, 250 ps of resolution |
External Test Clock (XClk) |
0 to 50MHz, TTL compatible input |
Pause Input |
External pause override, TTL compatible |
Trigger Input |
External trigger override, TTL compatible |
Run Status Output |
Run indicator status, TTL compatible |
VCC |
+5 VDC output |
Parametric Measurement Unit (PMU) |
|
Number of parametric measurement units |
32, one per channel |
Modes |
Force voltage, measure currentForce current, measure voltage |
Force voltage range |
- 10 volts to + 15 volts |
Force voltage accuracy |
+/- 25 mV, 25 volt range |
Force voltage and current resolution |
16 bits |
Force current range |
+/- 30 mA FS+/ 200 mA FS |
Force current accuracy |
+/- 100 uA, 30 mA range+/- 8 mA, 200 mA range |
Measure voltage range |
-10 to + 15 volts |
Measure voltage and current resolution |
16 bits |
Measure voltage accuracy |
+/- 15 mV, -9 to +13 volts range |
Measure current range |
+/- 20 mA FS+/- 200 mA FS |
Measure current accuracy |
+/- 100 uA (30 mA range)+/- 6 mA (200 mA range) |
Power Requirements |
|
3.3V Maximum current (mA): |
3.7A |
5V Maximum current (A): |
2A |
VCC, +18 volts, maximum current (A) |
6A |
VEE, -14 volts, maximum current (A) |
6A |
Environmental |
|
Operating Temp. |
0 to 50°C |
Storage Temperature |
-20 to 70°C |
Vibration |
5 g at 500 Hz |
Shock |
10 g for 6 ms, ½ sine |
Physical Properties |
|
Bus Interface |
PXI |
Dimensions |
6U, single slot PXI card |
Weight |
1.2 lbs (520 g) |
I/O Connections |
|
External Control and Status |
15 position, sub-D, female |
Digital I/O |
68 position SCSI III type |
External VCC and VEE |
9 position, sub –D, female |
GX5055 DIO Specifications