This section provides the GX2065 DMM specifications.
General Specifications |
|||||||||
Input |
Hi, Lo, Hi Sense, Lo Sense; floating and isolated from ground External trigger |
||||||||
Input Connectors |
(4) Banana,6 pin DIN |
||||||||
Format |
PXI, 3U single slot, hybrid slot compatible |
||||||||
DMM and Digitizer Measurement System Features |
|||||||||
Resolution |
22 bits (DMM), 16 bits (digitizer / DAQ) |
||||||||
DMM Sampling Rate(DCV, DCI & R measurements) |
Selectable PLC rate, 0.002 to 10; PLC can be set to 50, 60 or 400 Hz |
||||||||
DMM reading rate(DCV, DCI & 2WR (<10 Mohm)) |
5 readings/sec @ 6.5 digits, 60Hz (10 PLC), Auto Zero Off50 readings/sec @ 6.5 digits, 60 Hz (1 PLC) , Auto Zero Off500 readings/sec @ 5.5 digits, 60 Hz, (0.1 PLC) , Auto Zero Off, 1K Into Buffer2500 readings/sec @ 4.5 digits, 60 Hz, (0.01 PLC) , Auto Zero Off, 1K Into Buffer3500 readings/sec @ 3.5 digits, 60 Hz, (0.002 PLC) , Auto Zero Off, 1K Into Buffer |
||||||||
Digitizer (DAQ) Clock Rate |
Programmable to 3 MHzRange: (3MS/s) / N, N = 1 to 216Accuracy: 100 PPM |
||||||||
DAQ Measurement Functions |
AC and DC voltage and current measurements |
||||||||
Digitizer Memory |
8192 samples |
||||||||
DMM Memory |
1K samples |
||||||||
DMM Measurement Characteristics |
|||||||||
DCV Input range |
100 nV to 300 volts |
||||||||
ACV Input range |
3 uV to 425 V (peak); 300 VRMS, 10 Hz to 100 KHz, AC or DC coupled coupled |
||||||||
Crest Factor (ACV) |
No limitation as long as maximum input signal is below the peak value |
||||||||
DCV / ACV Input Impedance |
>10 G ohms (0.1, 1, and 10 volt DC ranges), 400 pF shunt capacitance10 M ohms for other AC/DC ranges, 400 pF shunt capacitance |
||||||||
Maximum Input (Volt Hertz) |
8 x 10e7 volt x Hz common mode input8 x 10e7 volt x Hz (across Hi or Lo input relative to earth ground) |
||||||||
Input Isolation |
CATII 300V |
||||||||
Input Overload Protection |
250V for current input,300V CATII for all other inputs |
||||||||
Noise Rejection |
DCV: 90 dB, NMRR; 140 dB CMRR15 readings/s, 1 PLC, 6.5 digit, 10 V rangeACV: 70dB, CMRR |
||||||||
DCI Input range |
10 nA to 2 amps |
||||||||
ACI Input range |
3 uA to 2 amps (peak), AC couples 10 Hz to 5 KHz |
||||||||
Crest Factor (AC current) |
No limitation as long as maximum input signal is below the maximum peak range value. |
||||||||
Burden Voltage (max) |
< 0.2V, 20 mA range<0.1 V, 100 mA range<0.5 V, 1 A range<1.0 V, 2A range |
||||||||
AC/DC Input Current Protection |
2A, 250V, fast blow, sand filled, 1.5kA breaking |
||||||||
Resistance Range |
0.1 m ohms to 100 M ohms |
||||||||
Resistance Measurement Configuration |
Selectable, 2-wire or 4-wire |
||||||||
Internal Temperature Measuring Accuracy |
+/- 2 C |
||||||||
Time Base Accuracy |
100 ppm |
||||||||
Triggering |
|||||||||
Trigger Source |
Function: Start measurementSource: PXI bus, software, continuous, external input (DIN connector), timer |
||||||||
Trigger Output Modes |
Functions: Start of measurement, end of measurementTrigger can be routed to the PXI bus or the DIN connector |
||||||||
Trigger Input Voltage Range |
3.3V CMOS, 5V Tolerant |
||||||||
Minimum Trigger Pulse Width |
50ns for PXI bus, 250us for external DIN input |
||||||||
Trigger Input Impedance |
4.75k ohms |
||||||||
Trigger Input Edge |
Selectable, positive or negative |
||||||||
AC Measurement Performance |
|||||||||
Reading Rate |
Signal Bandwidth |
||||||||
1 S/s |
4 Hz to 4 kHz |
||||||||
5 S/s |
20 Hz to 20 kHz (Default) |
||||||||
375 S/s |
300 Hz to 300 kHz |
||||||||
Voltage, Current, Resistance, and Frequency Measurement AccuracyAll accuracy specifications are relative to the calibration standard. |
|||||||||
DC Voltage |
|||||||||
Range |
Resolution |
Accuracy24 Hours 23°C ±5°(% of Reading) +(% of FS) |
Accuracy90 Days 23°C ±5°(% of Reading )+(% of FS) |
Accuracy1 Year 23°C ±5°(% of Reading) +(% of FS) |
|||||
100mV |
100nV |
0.0030 + 0.0040 |
0.0040 + 0.0045 |
0.0045 + 0.0045 |
|||||
1V |
1 µV |
0.0030 + 0.0007 |
0.0040 + 0.0008 |
0.0045 + 0.0008 |
|||||
10V |
10 µV |
0.0010 + 0.0004 |
0.0025 + 0.0005 |
0.0030 + 0.0005 |
|||||
100V |
100 µV |
0.0030 + 0.0006 |
0.0050 + 0.0009 |
0.0060 + 0.0009 |
|||||
300V |
1 mV |
0.0030 + 0.0020 |
0.0045 + 0.0030 |
0.0060 + 0.0030 |
|||||
Notes:DC measurements @ 10 PLC or 1 PLC with digital filtering |
|||||||||
DC Current |
|||||||||
Range |
Resolution |
Accuracy24 Hours 23°C ±5°(% of Reading) +(% of FS) |
Accuracy90 Days 23°C ±5°(% of Reading)+(% of FS) |
Accuracy1 Year 23°C ±5°(% of Reading) +(% of FS) |
|||||
20 mA |
10 nA |
0.0060 + 0.0030 |
0.0030 + 0.0080 |
0.0500 + 0.0080 |
|||||
100 mA |
100 nA |
0.0100 + 0.0300 |
0.0300 + 0.0800 |
0.0500 + 0.0800 |
|||||
1A |
1 uA |
0.0200 + 0.0030 |
0.0500 + 0.0080 |
0.0800 + 0.0080 |
|||||
2A |
10 uA |
0.1000 + 0.0035 |
0.1200 + 0.0060 |
0.1200 + 0.0060 |
|||||
DC Current Measurement, DAQ Mode |
|||||||||
Range |
Resolution |
Accuracy24 Hours 23°C ±5°(% of Reading) +(% of FS) |
Accuracy90 Days 23°C ±5°(% of Reading) +(% of FS) |
Accuracy1 Year 23°C ±5°(% of Reading) +(% of FS) |
|||||
20mA |
4 uA |
0.030 + 1.00 |
0.030 + 1.00 |
0.030 + 1.00 |
|||||
100mA |
40 uA |
0.010 + 1.00 |
0.010 + 1.00 |
0.010 + 1.00 |
|||||
1A |
40 uA |
0.040 + 0.150 |
0.040 + 0.150 |
0.040 + 0.150 |
|||||
2A |
80 uA |
0.040 + 0.150 |
0.040 + 0.150 |
0.040 + 0.150 |
|||||
AC Volts (RMS), AC Coupled, DAQ Mode |
|||||||||
Range (RMS) |
Resolution |
Frequency |
Accuracy24 Hours 23°C ±5°(% of Reading) +(% of FS) |
Accuracy90 Days 23°C ±5°(% of Reading) +(% of FS) |
Accuracy1 Year 23°C ±5°(% of Reading) +(% of FS) |
||||
50mV |
2 uV |
3 Hz - 10Hz |
0.5 + 0.28 |
0.5 + 0.28 |
0.5 + 0.28 |
||||
>10 Hz 20 KHz |
0.2 + 0.28 |
0.2 + 0.28 |
0.2 + 0.28 |
||||||
> 20 KHz 50 KHz |
0.26 + 0.3 |
0.26 + 0.3 |
0.26 + 0.3 |
||||||
> 50 KHz 100 KHz |
0.75 + 0.33 |
0.75 + 0.33 |
0.75 + 0.33 |
||||||
> 100 KHz 300 KHz |
4.15 + 0.75 |
4.15 + 0.75 |
4.15 + 0.75 |
||||||
0.5 V5 V50 V300 V |
20 uV200 uV2 mV30 mV |
3 Hz - 10Hz |
0.35 + 0.03 |
0.35 + 0.03 |
0.35 + 0.03 |
||||
> 10 Hz 20 KHz |
0.05 + 0.03 |
0.05 + 0.03 |
0.06 + 0.03 |
||||||
> 20 KHz 50 KHz |
0.11 + 0.05 |
0.11 + 0.05 |
0.12 + 0.05 |
||||||
> 50 KHz 100 KHz |
0.60 + 0.08 |
0.60 + 0.08 |
0.60 + 0.08 |
||||||
> 100 KHz 300 KHz |
4.0 + 0.5 |
4.0 + 0.5 |
4.0 + 0.5 |
||||||
AC Current (RMS), AC Coupled, DAQ Mode |
|||||||||
Range (RMS) |
Resolution |
Frequency |
Accuracy24 Hours 23°C ±5°(% of Reading) +(% of FS) |
Accuracy90 Days 23°C ±5°(% of Reading) +(% of FS) |
Accuracy1 Year 23°C ±5°(% of Reading) +(% of FS) |
||||
0.5 amp |
40 uA |
3 Hz - 10 Hz |
0.30 + 0.04 |
0.30 + 0.04 |
0.30 + 0.04 |
||||
> 10 Hz 3 KHz |
0.10 + 0.04 |
0.10 + 0.04 |
0.10 + 0.04 |
||||||
> 3 KHz - 5 KHz |
0.14 + 0.04 |
0.14 + 0.04 |
0.14 + 0.04 |
||||||
1.0 amp |
80 uA |
3 Hz - 10 Hz |
0.35 + 0.09 |
0.35 + 0.09 |
0.35 + 0.09 |
||||
> 10 Hz 3 KHz |
0.15 + 0.09 |
0.15 + 0.09 |
0.15 + 0.09 |
||||||
> 3 KHz - 5 KHz |
0.18 + 0.09 |
0.18 + 0.09 |
0.18 + 0.09 |
||||||
Resistance |
|||||||||
Range (ohms) |
Open Circuit Voltage & Test Current |
Resolution |
Accuracy24 Hours 23°C ±5°(% of Reading) +(% of FS) |
Accuracy90 Days 23°C ±5°(% of Reading) +(% of FS) |
Accuracy1 Year 23°C ±5°(% of Reading) +(% of FS) |
||||
100 |
6.9V, 1 mA |
100 µ ohms |
0.0020 + 0.0060 |
0.0080 + 0.0060 |
0.0100 + 0.0060 |
||||
1000 |
6.9V, 1 mA |
1 m ohm |
0.0025 + 0.0006 |
0.0085 + 0.0020 |
0.0105 + 0.0006 |
||||
10K |
6.9V, 100 uA |
10 m ohm |
0.0020 + 0.0006 |
0.0080 + 0.0020 |
0.0100 + 0.0006 |
||||
100K |
12.8V, 10 uA |
100 m ohms |
0.0030 + 0.0006 |
0.0090 + 0.0010 |
0.0110 + 0.0010 |
||||
1 M |
12.8V, 1 uA |
1 ohm |
0.0020 + 0.0006 |
0.0020 + 0.0010 |
0.0100 + 0.0010 |
||||
10 M |
7V, 0.7 uA //10 M ohm |
10 ohms |
0.0150 + 0.0006 |
0.0200 + 0.0010 |
0.0400 + 0.0010 |
||||
100 M |
7V, 0.7 uA // 10 M ohm |
100 ohms |
0.0800 + 0.0030 |
0.2000 + 0.0030 |
0.2000 + 0.0030 |
||||
DC Voltage Measurement, DAQ Mode |
|||||||||
Range |
Resolution |
Accuracy24 Hours 23°C ±5°(% of Reading) +(% of FS) |
Accuracy90 Days 23°C ±5°(% of Reading) +(% of FS) |
Accuracy1 Year 23°C ±5°(% of Reading) +(% of FS) |
|||||
100 mV |
4 uV |
0.06 + 0.06 |
0.06 + 0.06 |
0.06 + 0.06 |
|||||
1 V |
40 uV |
0.06 + 0.03 |
0.06 + 0.03 |
0.06 + 0.03 |
|||||
10 V |
400 uV |
0.06 + 0.03 |
0.06 + 0.03 |
0.06 + 0.03 |
|||||
100 V |
4 mV |
0.06 + 0.03 |
0.06 + 0.03 |
0.06 + 0.03 |
|||||
300 V |
40 mV |
0.06 + 0.10 |
0.06 + 0.10 |
0.06 + 0.10 |
|||||
Diode Voltage Measurement |
|||||||||
Range |
Voltage Range |
Current Source Accuracy |
Accuracy (Nominal)24 Hours 23°C ±5°(% of Reading)+(% of FS) |
Accuracy(Nominal)90 Days 23°C ±5°(% of Reading) +(% of FS) |
Accuracy(Nominal)1 Year 23°C ±5°(% of Reading) +(% of FS) |
||||
10 uA |
0-12 V |
5% |
0.0010 + 0.0004 |
0.0025 + 0.0005 |
0.0030 + 0.0005 |
||||
100 uA |
0-6.5 V |
5% |
0.0010 + 0.0004 |
0.0025 + 0.0005 |
0.0030 + 0.0005 |
||||
1 mA |
0-6.5 V |
5% |
0.0010 + 0.0004 |
0.0025 + 0.0005 |
0.0030 + 0.0005 |
||||
Frequency Measurement |
|||||||||
Frequency Range |
1 Hz to 500 KHz |
||||||||
Input Voltage* |
20 mV to 300 V |
||||||||
Resolution (offset ppm) |
0.33 (1 second gate time)3.33 (100 mSec gate time)33.3 (10 mSec gate time) |
||||||||
Accuracy |
100 ppm of reading + offset ppm |
||||||||
* Input amplitude must be at least 20% of FS and input amplitude must not exceed specified volt hertz product. |
|||||||||
Environmental and Physical Specifications |
|||||||||
Safety |
Complies with IEC 61010-1, CAT II 300V, pollution degree 2 |
||||||||
EMC |
Complies with EN61326-1 |
||||||||
Calibration |
Calibration is performed at the factory using NIST traceable instrumentation. All calibration constants are stored on-board in non-volatile EEROM. Calibration can be performed by any calibration laboratory with the appropriate equipment. |
||||||||
Temperature Range |
Operating: 0 to + 50°CExtended operating range: -20°C to +70°CStorage: -20°C to +70°C storage |
||||||||
Relative Humidity |
Operating: 80% at 40°CStorage to 95% at40°C |
||||||||
Power (max) |
+5VDC, 2.3 A+3.3 VDC, 255 mA+12 VDC, 16 mA-12 VDC, 25 mA |
||||||||
Connectors |
(4) Banana jacks:Hi: Voltage, 2W ohmsLo: Voltage, Current, 2W ohmsSense Hi: Current, 4W ohmsSense Lo: 4W ohms6-pin DIN: Trigger in, Trigger out, Trigger Gnd |
||||||||
Note: Specifications are subject to change.