Appendix A - Specifications

This section provides the GX2065 DMM specifications.

General Specifications
Input
Hi, Lo, Hi Sense, Lo Sense; floating and isolated from ground External trigger
Input Connectors
(4) Banana,
6 pin DIN
Format
PXI, 3U single slot, hybrid slot compatible
DMM and Digitizer Measurement System Features
Resolution
22 bits (DMM), 16 bits (digitizer / DAQ)
DMM Sampling Rate
(DCV, DCI & R measurements)
Selectable PLC rate, 0.002 to 10; PLC can be set to 50, 60 or 400 Hz
DMM reading rate
(DCV, DCI & 2WR (<10 Mohm))
5 readings/sec @ 6.5 digits, 60Hz (10 PLC), Auto Zero Off
50 readings/sec @ 6.5 digits, 60 Hz (1 PLC) , Auto Zero Off
500 readings/sec @ 5.5 digits, 60 Hz, (0.1 PLC) , Auto Zero Off, 1K Into Buffer
2500 readings/sec @ 4.5 digits, 60 Hz, (0.01 PLC) , Auto Zero Off, 1K Into Buffer
3500 readings/sec @ 3.5 digits, 60 Hz, (0.002 PLC) , Auto Zero Off, 1K Into Buffer
Digitizer (DAQ) Clock Rate
Programmable to 3 MHz
Range: (3MS/s) / N, N = 1 to 216
Accuracy: 100 PPM
DAQ Measurement Functions
AC and DC voltage and current measurements
Digitizer Memory
8192 samples
DMM Memory
1K samples
DMM Measurement Characteristics
DCV Input range
100 nV to 300 volts
ACV Input range
3 uV to 425 V (peak); 300 VRMS, 10 Hz to 100 KHz, AC or DC coupled coupled
Crest Factor (ACV)
No limitation as long as maximum input signal is below the peak value
DCV / ACV Input Impedance
>10 G ohms (0.1, 1, and 10 volt DC ranges), 400 pF shunt capacitance
10 M ohms for other AC/DC ranges, 400 pF shunt capacitance
Maximum Input (Volt Hertz)
8 x 10e7 volt x Hz common mode input
8 x 10e7 volt x Hz (across Hi or Lo input relative to earth ground)
Input Isolation
CATII 300V
Input Overload Protection
250V for current input,
300V CATII for all other inputs
Noise Rejection
DCV: 90 dB, NMRR; 140 dB CMRR
15 readings/s, 1 PLC, 6.5 digit, 10 V range
ACV: 70dB, CMRR
DCI Input range
10 nA to 2 amps
ACI Input range
3 uA to 2 amps (peak), AC couples 10 Hz to 5 KHz
Crest Factor (AC current)
No limitation as long as maximum input signal is below the maximum peak range value.
Burden Voltage (max)
< 0.2V, 20 mA range
<0.1 V, 100 mA range
<0.5 V, 1 A range
<1.0 V, 2A range
AC/DC Input Current Protection
2A, 250V, fast blow, sand filled, 1.5kA breaking
Resistance Range
0.1 m ohms to 100 M ohms
Resistance Measurement Configuration
Selectable, 2-wire or 4-wire
Internal Temperature Measuring Accuracy
+/- 2 C
Time Base Accuracy
100 ppm
Triggering
Trigger Source
Function: Start measurement
Source: PXI bus, software, continuous, external input (DIN connector), timer
Trigger Output Modes
Functions: Start of measurement, end of measurement
Trigger can be routed to the PXI bus or the DIN connector
Trigger Input Voltage Range
3.3V CMOS, 5V Tolerant
Minimum  Trigger Pulse Width
50ns for PXI bus, 250us for external DIN input
Trigger Input Impedance
4.75k ohms
Trigger Input Edge
Selectable, positive or negative
AC Measurement Performance
Reading Rate
Signal Bandwidth
1 S/s
4 Hz to 4 kHz
5 S/s
20 Hz to 20 kHz (Default)
375 S/s
300 Hz to 300 kHz
Voltage, Current,  Resistance, and Frequency Measurement Accuracy
All accuracy specifications are relative to the calibration standard.
DC Voltage
Range
Resolution
Accuracy
24 Hours 23°C ±5°
(% of Reading) +
(% of FS)
Accuracy
90 Days 23°C ±5°
(% of Reading )+
(% of FS)
Accuracy
1 Year 23°C ±5°
(% of Reading) +
(% of FS)
100mV
100nV
0.0030 + 0.0040
0.0040 + 0.0045
0.0045 + 0.0045
1V
1 µV
0.0030 + 0.0007
0.0040 + 0.0008
0.0045 + 0.0008
10V
10 µV
0.0010 + 0.0004
0.0025 + 0.0005
0.0030 + 0.0005
100V
100 µV
0.0030 + 0.0006
0.0050 + 0.0009
0.0060 + 0.0009
300V
1 mV
0.0030 + 0.0020
0.0045 + 0.0030
0.0060 + 0.0030
Notes:
DC measurements @ 10 PLC or 1 PLC with digital filtering
DC Current
Range
Resolution
Accuracy
24 Hours 23°C ±5°
(% of Reading) +
(% of FS)
Accuracy
90 Days 23°C ±5°
(% of Reading)+
(% of FS)
Accuracy
1 Year 23°C ±5°
(% of Reading) +
(% of FS)
20 mA
10 nA
0.0060 + 0.0030
0.0030 + 0.0080
0.0500 + 0.0080
100 mA
100 nA
0.0100 + 0.0300
0.0300 + 0.0800
0.0500 + 0.0800
1A
1 uA
0.0200 + 0.0030
0.0500 + 0.0080
0.0800 + 0.0080
2A
10 uA
0.1000 + 0.0035
0.1200 + 0.0060
0.1200 + 0.0060
DC Current Measurement, DAQ Mode
Range
Resolution
Accuracy
24 Hours 23°C ±5°
(% of Reading) +
(% of FS)
Accuracy
90 Days 23°C ±5°
(% of Reading) +
(% of FS)
Accuracy
1 Year 23°C ±5°
(% of Reading) +
(% of FS)
20mA
4 uA
0.030 + 1.00
0.030 + 1.00
0.030 + 1.00
100mA
40 uA
0.010 + 1.00
0.010 + 1.00
0.010 + 1.00
1A
40 uA
0.040 +  0.150
0.040 +  0.150
0.040 +  0.150
2A
80 uA
0.040 +  0.150
0.040 +  0.150
0.040 +  0.150
AC Volts (RMS), AC Coupled, DAQ Mode
Range (RMS)
Resolution
Frequency
Accuracy
24 Hours 23°C ±5°
(% of Reading) +
(% of FS)
Accuracy
90 Days 23°C ±5°
(% of Reading) +
(% of FS)
Accuracy
1 Year 23°C ±5°
(% of Reading) +
(% of FS)
50mV
 
2 uV
 
3 Hz - 10Hz
0.5 + 0.28
0.5 + 0.28
0.5 + 0.28
>10 Hz 20 KHz
0.2 + 0.28
0.2 + 0.28
0.2 + 0.28
> 20 KHz 50 KHz
0.26 + 0.3
0.26 + 0.3
0.26 + 0.3
> 50 KHz 100 KHz
0.75 + 0.33
0.75 + 0.33
0.75 + 0.33
> 100 KHz 300 KHz
4.15 + 0.75
4.15 + 0.75
4.15 + 0.75
0.5 V
5 V
50 V
300 V
20 uV
200 uV
2 mV
30 mV
3 Hz - 10Hz
0.35 + 0.03
0.35 + 0.03
0.35 + 0.03
> 10 Hz 20 KHz
0.05 + 0.03
0.05 + 0.03
0.06 + 0.03
> 20 KHz 50 KHz
0.11 + 0.05
0.11 + 0.05
0.12 + 0.05
> 50 KHz 100 KHz
0.60 + 0.08
0.60 + 0.08
0.60 + 0.08
> 100 KHz 300 KHz
4.0 + 0.5
4.0 + 0.5
4.0 + 0.5
AC Current (RMS), AC Coupled, DAQ Mode
Range (RMS)
Resolution
Frequency
Accuracy
24 Hours 23°C ±5°
(% of Reading) +
(% of FS)
Accuracy
90 Days 23°C ±5°
(% of Reading) +
(% of FS)
Accuracy
1 Year 23°C ±5°
(% of Reading) +
(% of FS)
0.5 amp
40 uA
3 Hz - 10 Hz
0.30 + 0.04
0.30 + 0.04
0.30 + 0.04
> 10 Hz 3 KHz
0.10 + 0.04
0.10 + 0.04
0.10 + 0.04
> 3 KHz - 5 KHz
0.14 + 0.04
0.14 + 0.04
0.14 + 0.04
1.0 amp
 
80 uA
3 Hz - 10 Hz
0.35 + 0.09
0.35 + 0.09
0.35 + 0.09
> 10 Hz 3 KHz
0.15 + 0.09
0.15 + 0.09
0.15 + 0.09
> 3 KHz - 5 KHz
0.18 + 0.09
0.18 + 0.09
0.18 + 0.09
Resistance
Range (ohms)
Open Circuit Voltage & Test Current
Resolution
Accuracy
24 Hours 23°C ±5°
(% of Reading) +
(% of FS)
Accuracy
90 Days 23°C ±5°
(% of Reading) +
(% of FS)
Accuracy
1 Year 23°C ±5°
(% of Reading) +
(% of FS)
100
6.9V, 1 mA
100 µ ohms
0.0020 + 0.0060
0.0080 + 0.0060
0.0100 + 0.0060
1000
6.9V, 1 mA
1 m ohm
0.0025 + 0.0006
0.0085 + 0.0020
0.0105 + 0.0006
10K
6.9V, 100 uA
10 m ohm
0.0020 + 0.0006
0.0080 + 0.0020
0.0100 + 0.0006
100K
12.8V, 10 uA
100 m ohms
0.0030 + 0.0006
0.0090 + 0.0010
0.0110 + 0.0010
1 M
12.8V, 1 uA
1 ohm
0.0020 + 0.0006
0.0020 + 0.0010
0.0100 + 0.0010
10 M
7V, 0.7 uA //10 M ohm
10 ohms
0.0150 + 0.0006
0.0200 + 0.0010
0.0400 + 0.0010
100 M
7V, 0.7 uA // 10 M ohm
100 ohms
0.0800 + 0.0030
0.2000 + 0.0030
0.2000 + 0.0030
DC Voltage Measurement, DAQ Mode
Range
Resolution
Accuracy
24 Hours 23°C ±5°
(% of Reading) +
(% of FS)
Accuracy
90 Days 23°C ±5°
(% of Reading) +
(% of FS)
Accuracy
1 Year 23°C ±5°
(% of Reading) +
(% of FS)
100 mV
4 uV
0.06 + 0.06
0.06 + 0.06
0.06 + 0.06
1 V
40 uV
0.06 + 0.03
0.06 + 0.03
0.06 + 0.03
10 V
400 uV
0.06 + 0.03
0.06 + 0.03
0.06 + 0.03
100 V
4 mV
0.06 + 0.03
0.06 + 0.03
0.06 + 0.03
300 V
40 mV
0.06 + 0.10
0.06 + 0.10
0.06 + 0.10
Diode Voltage Measurement
Range
 
Voltage Range
Current Source Accuracy
Accuracy (Nominal)
24 Hours 23°C ±5°
(% of Reading)+
(% of FS)
Accuracy
(Nominal)
90 Days 23°C ±5°
(% of Reading) +
(% of FS)
Accuracy
(Nominal)
1 Year 23°C ±5°
(% of Reading) +
(% of FS)
10 uA
0-12 V
5%
0.0010 + 0.0004
0.0025 + 0.0005
0.0030 + 0.0005
100 uA
0-6.5 V
5%
0.0010 + 0.0004
0.0025 + 0.0005
0.0030 + 0.0005
1 mA
0-6.5 V
5%
0.0010 + 0.0004
0.0025 + 0.0005
0.0030 + 0.0005
Frequency Measurement
Frequency Range
1 Hz to 500 KHz
Input Voltage*
20 mV to 300 V
Resolution (offset ppm)
0.33 (1 second gate time)
3.33 (100 mSec gate time)
33.3 (10 mSec gate time)
Accuracy
100 ppm of reading + offset ppm
* Input amplitude must be at least 20% of FS and input amplitude must not exceed specified volt hertz product.
Environmental and Physical Specifications
Safety
Complies with IEC 61010-1, CAT II 300V, pollution degree 2
EMC
Complies with EN61326-1
Calibration
Calibration is performed at the factory using NIST traceable instrumentation. All calibration constants are stored on-board in non-volatile EEROM. Calibration can be performed by any calibration laboratory with the appropriate equipment.
Temperature Range
Operating: 0 to + 50°C
Extended operating range: -20°C to +70°C
Storage: -20°C to +70°C storage
Relative Humidity
Operating: 80% at 40°C
Storage to 95% at40°C
Power (max)
+5VDC, 2.3 A
+3.3 VDC, 255 mA
+12 VDC, 16 mA
-12 VDC, 25 mA
Connectors
(4) Banana jacks:
Hi: Voltage, 2W ohms
Lo: Voltage, Current, 2W ohms
Sense Hi: Current, 4W ohms
Sense Lo: 4W ohms
6-pin DIN: Trigger in, Trigger out, Trigger Gnd

 

Note: Specifications are subject to change.