The following table outlines the specifications of the GX1034. All accuracies are based on values measured with a NIST traceable 8.5-digit DMM and stored in the GX1034 card’s EEPROM.
4-wire Resistance |
|
Value |
1 Ω, nominal |
AccuracyStability |
+/- 0.06%3 ppm / °C |
Value |
10 Ω nominal |
AccuracyStability |
+/- 0.012%0.2 ppm / °C |
Values |
100 Ω nominal |
AccuracyStability |
+/- 0.01%0.2 ppm / °C |
2-wire Resistance |
|
Value |
1K Ω nominal |
AccuracyStability |
+/- 0.004%0.2 ppm / °C |
Value |
10K Ω nominal |
AccuracyStability |
+/- 0.004%0.2 ppm / °C |
Value |
100K Ω nominal |
AccuracyStability |
+/- 0.015%5 ppm /° C |
Value |
1M Ω nominal |
AccuracyStability |
+/- 0.0275%3 ppm / °C |
Value |
10M Ω nominal |
AccuracyStability |
+/- 0.1%2 ppm / °C |
DC Voltage Source |
|
Values |
+/ - 9.0V, +0.9V, +0.09V, +0.009V, nominal |
Accuracy |
±9.0V: ±0.005% ±0.9V: ±0.005% ±0.09V: ±0.016% ±0.009V: ±0.1% |
Temperature Coefficient |
0.6 ppm / °C |
Long Term Drift |
6 ppm for 1000 operating hours |
Output Divider |
Resistive series network:10 KΩ, 1 KΩ, 100 Ω, 10 Ω |
AC Voltages (RMS), Sine Waveform |
|
Values: |
3.5VAC, 0.35VAC, 0.035VAC, 0.0035VAC, nominal50 Hz, 60 Hz, 100 Hz frequencies |
Accuracy at: 50 Hz, 60 Hz, 100 Hz |
3.5VAC: ±0.6% 0.35VAC: ±0.6% 0.035VAC: ±0.6% 0.0035VAC: ±4% |
Accuracy at: 400 Hz, 1KHz |
3.5VAC: ±0.6% 0.35VAC: ±0.6% 0.035VAC: ±0.6% 0.0035VAC: ±4.75% |
Accuracy at: 10 KHz |
3.5VAC: ±0.6% 0.35VAC: ±0.6% 0.035VAC: ±0.6% 0.0035VAC: ±5% |
Accuracy at: 100 KHz |
3.5VAC: ±4% 0.35VAC: ±4% 0.035VAC: ±4% 0.0035VAC: ±9.5% |
Temperature Coefficient |
3 ppm / °C, (voltage reference for D/A) |
Long Term Drift |
6 ppm for 1000 operating hours |
Frequency Accuracy |
+/- 0.1 ppm |
Output Divider |
Resistive series network:10 KΩ, 1 KΩ, 100 Ω, 10 Ω |
10 MHz Frequency Output Reference |
|
Source |
10 MHz crystal oscillator |
Initial accuracy |
±1 ppm |
Drift |
100 ppb / °C , max |
10 MHz Output |
Front panel BNC, TTL compatible, 50-ohm load PXI connector, PXI 10 MHz clock |
10 MHz Divider Output |
24-bit, programmable |
DC Current Source |
|
Range |
0-19.9mA, 16-bit resolution |
Accuracy |
+/- .025% of programmed value |
Compliance |
0 – 12.5 volts |
Voltage Measurement (ADC) |
|
Resolution |
16-bit |
Mode |
Differential |
Input impedance |
20 K Ω nominal |
Range |
–10V to +10V |
Accuracy |
± 0.025V of reading |
Power Requirements |
|
3.3V |
500 mA |
5V |
1 A |
12V |
500 mA |
Physical |
|
Size |
3U, single slot, hybrid PXI slot compatible |
I/O Connectors |
DB-9, female; BNC for 10 MHz reference & divider output |
Weight |
210 grams |
Operating Temp |
0 to 50°C |
Storage Temp |
0 to 70°C |
Humidity |
10% - 80% non-condensing |
Safety |
EN61010-1:2001 |
CE Labeled |
Yes, EN61000-6-1:2001:EN55011:1998 |
Calibration Interval |
1 year |