Specifications

The following table outlines the specifications of the GX1838.

Output channels
8
Voltage Rails
3
Voltage source
3
Output Voltage Range (all voltage sources)
-10 V to +32 VDC (Standard version)
-20 V to +20 VDC (GX1838-20 version)
Resolution
14-bit
Accuracy
± 14mV
(with 1000 Ohm load)
Slew Rate
6 V/µs
Max Load (min. resistance) per channel
50Ω
Max. Current per channel
500mA
Max. Current per rail
500mA
Max. Current per GX1838
500mA
Output Noise
< 4mV rms, 1K Ohm load, 20 - 300 KHz measurement BW
Output Protection
Short Circuit (Ground)
Power ON State
All Channels / Rails open
Operating Temperature
0 to+55°C
Storage Temperature
-20 to+70°C
Size
3U PXI, hybrid slot compatible
Weight
14 oz.

 

Power Requirements

3.3 VDC Current
2.1A max
5    VDC Current
2.9A max

Calibration

Calibration Interval

1 year