GX3104, GX3116e
Sets the specified channel’s measurement parameters.
GxSmuMeasureSetParamters (nHandle, nApertureType, dApertureTime, pnStatus)
Name |
Type |
Description |
nHandle |
SHORT |
Handle to a GX3104 or GX3116e board. |
nApertureTime |
SHORT |
0. GXSMU_MEASURE_IN_READINGS_PER_SEC: Set the measurement in reading per second1. GXSMU_MEASURE_IN_PLC: Set the measurement in Power Line Cycles |
dAperture |
DOUBLE |
Value depends on the nApertureTimeType parameter settings. Capture aperture in terms of Power Line Cycles or in seconds. |
pnStatus |
PSHORT |
Returned status: 0 on success, negative number on failure. |
The aperture can be expressed in terms of Power Line Cycles as set by the GxSmuMeasureSetPowerLineFrequency function, or in seconds. The Aperture can be a fraction of a line cycle. For example, the aperture can be set to 0.1 Line Cycles which would represent approximately 0.00167 seconds if the Line Frequency was set to 60Hz.
The following example sets measurement parameters:
SHORT nHandle, nStatus, nApertureType;
DOUBLE dApertureL
GxSmuMeasureSetParamters (nHandle, GXSMU_MEASURE_IN_READINGS_PER_SEC, 1, &nStatus);
GxSmuMeasureGetParamters (nHandle, &nApertureType, &dApertureL, &nStatus);
GxSmuMeasureSetSenseMode, GxSmuMeasureSetPowerLineFrequency, GxSmuMeasure, GxSmuMeasureGetParamters, GxSmuGetErrorString