GX3104, GX3116e
Returns the specified channel’s current settings when in constant current mode.
GxSmuSourceGetCurrent (nHandle, nChannel, pdCurrent, pnStatus)
Name |
Type |
Description |
nHandle |
SHORT |
Handle to a GX3104 or GX3116e board. |
nChannel |
SHORT |
GX3104: Specifies the channel number (0-3): GX3104_CHANNEL_1 to GX3104_CHANNEL_4 GX3116e: Specifies channel number (0-15) is GX3116_CHANNEL_1 to GX3116_CHANNEL_16. |
pdCurrent |
PDOUBLE |
Returns the specified channel’s current settings, the range for each current range is as follows:GX3104:0. GX3104_SOURCE_CURRENT_RANGE_1000MA: Source Current Range +/-1000mA. Supported by channel 1 only, channels 2-4 can be set up to +/-500mA.1. GX3104_SOURCE_CURRENT_RANGE_250MA: Source Current Range +/-250mA2. GX3104_SOURCE_CURRENT_RANGE_25MA: Source Current Range +/-25mA (default after reset).3. GX3104_SOURCE_CURRENT_RANGE_2500UA: Source Current Range +/-2.5mA4. GX3104_SOURCE_CURRENT_RANGE_250UA: Source Current Range +/-250uA5. GX3104_SOURCE_CURRENT_RANGE_25UA: Source Current Range +/-25uA6. GX3104_SOURCE_CURRENT_RANGE_2500NA: Source Current Range -2.5uA to -1uA, and 1uA to 2.5uA.GX3116e:0. GX3116_SOURCE_CURRENT_RANGE_512MA: Source Current Range -512mA to +512mA1. GX3116_SOURCE_CURRENT_RANGE_25_POINT_6MA: Source Current Range -25.6mA to +25.6mA (default)2. GX3116_SOURCE_CURRENT_RANGE_2_POINT_56MA: Source Current Range -2.56mA to +2.56mA |
pnStatus |
PSHORT |
Returned status: 0 on success, negative number on failure. |
The function returns the specified channel’s current settings. The returned current value is in mA.
The following example returns channel 1 current settings
SHORT nStatus;
DOUBLE dCurrent;
GxSmuSourceGetCurrent (nHandle, GX3104_CHANNEL_1, &dCurrent, &nStatus);
GxSmuSourceSetCurrent, GxSmuSourceSetCurrentLimit, GxSmuSourceSetCurrentClamps, GxSmuSourceSetCurrentRange, GxSmuSourceSetMode, GxSmuSourceSetOutputState, GxSmuSourceSetRangeMode, GxSmuSourceSetVoltage, GxSmuSourceSetVoltageLimit, GxSmuGetErrorString