The following table details the specifications of the GX3116e.
Source and Measure Specifications per channel (total of 16 channels) |
|||||||||||||||
Voltage Source(All Current ranges) |
Resolution: 16-bitCurrent Range 512 mA: Output Voltage range: 0.0 VDC to +9.0 VDCAccuracy: ±15mVCurrent Range 25.6 mA and 2.56 mA: Output Voltage range: -2.0 VDC to +10.5 VDCAccuracy: ±15mV between -1.75 VDC and +10.5 VDC, otherwise ±25mVCurrent Range 2.56 mA: Output Voltage range: -2.0 VDC to +10.5 VDC Accuracy: ±15mV between -1.75 VDC and +10.5 VDC, otherwise ±25mV |
||||||||||||||
Voltage Source Current Clamp Source and Sink |
Resolution: 16-bitCurrent Range 512 mA: 75 mA to 650 mA: Accuracy: ± (20 mA +1.0% of FS range)Current Range 25.6 mA: 7.5 mA to 30 mA: Accuracy: ±(1.5 mA + 1.0% of FS range)Current Range 2.56 mA: 1 mA to 3.0 mA: Accuracy: ±(0.10 mA +5.0% of FS range) |
||||||||||||||
Output Voltage Ripple and Noise |
< 10 mV p-p, 20 MHz BW, full load |
||||||||||||||
Current Source |
Resolution: 16-bitCurrent Range +/-512 mA: Accuracy: ±0.5% of FS rangeCurrent Range +/-25.6 mA: Accuracy: ±0.3% of FS rangeCurrent Range +/-2.56 mA: Accuracy: ±0.4% of FS range |
||||||||||||||
Source Noise |
20 uV RMS (0.1 Hz to 10 Hz)25 mV RMS (20 Hz to 20 MHz) |
||||||||||||||
Settling Time |
Settles within 0.1% of specified value in <= 1.2mSec |
||||||||||||||
Transient Response |
Settles to ±20 mV of programmed value after 80% load change, 500 us typical |
||||||||||||||
Voltage Measurement(All Current ranges) |
Range: -2.0 VDC to +10.5 VDCAccuracy:+/-3mV when measuring voltage -1. 25V to
10.5V,
|
||||||||||||||
Conversion (mSec) |
Effective number of Digits |
4.0 |
6.02 |
1.5 |
5.80 |
1.0 |
5.63 |
0.8 |
5.57 |
0.5 |
5.42 |
0.4 |
5.33 |
Note: For applications that require maximum output power from the GX3116e (60 watts), this module will exceeds the power capabilities of a standard PXI chassis (25 watts). If required, a high-power chassis such as the GX7205 or GX7305 can be used which offers 60 watts of power per slot.