The following table outlines the specifications of the GX6021.
Maximum switchable voltage(signal to ground ) |
170VDC or peak AC, resistive load |
Maximum switchable current |
0.5A DC Resistive load |
Maximum carry current |
0.5A DC Resistive load |
Relay Contact Resistance |
<0.2 W |
RF specifications |
|
Impedance |
50 Ohm |
Bandwidth (-3 dB ) |
>300MHz |
Insertion Loss |
<0.7 dB at 100MHz |
Crosstalk across groups (50- ohm terminator) |
10 MHz < -40 dB100 MHz <-40 dB200 MHz <-35 dB |
Isolation between channels (50-ohm terminator) |
10 MHz < -40 dB100 MHz <-40 dB300 MHz <-35 dB |
VSWR (50-ohm terminator) |
1.3 : 1 at 250 MHz |
Temperature:Operating:Storage: |
0 to+40°C-10 to+70°C |
Vibration |
5g at 500Hz |
Shock ½ Sine |
5g for 6mS |
3.3 V Power |
160mA |
5V Typical Power |
20mA per channel |
5V Maximum Power |
1.25A max |
Size |
3U PXI |
Weight |
GX6021: 570g |